Opt Express ; 32(7): 12774-12782, 2024 Mar 25.
Artigo em Inglês | MEDLINE | ID: mdl-38571090

RESUMO

We demonstrate a high dynamic range (DR) Fourier-transform-based terahertz (THz) spectrometer by combining a THz photomultiplier tube (PMT) with a metasurface and a conventional Michelson interferometer. Because the THz-PMT response depends on the incident electric-field strength following the Fowler-Nordheim equation, we can directly obtain an electric field interferogram without any synchronized optical probe pulse in contrast to conventional THz-time-domain-spectroscopy (THz-TDS). The DR of the corresponding power spectrum using the proposed method was 4.6 × 105 without the use of a lock-in amplifier. The complex refractive index of a quartz glass plate obtained using the proposed method was in good agreement with the results of conventional THz-TDS.